PEI Protocol Specification

Physical Execution Interface (PEI) — the async protocol for connecting AI orchestrators to physical robot hardware in science labs.

Version: 0.1.0 | Status: Draft


1. Layer Model

PEI organizes robot actions into four layers:

Layer Name Actions Description
0 Motion move, grip, release, home Raw kinematic primitives
1 Lab-Ops pipette, transfer_labware, dispense Domain-specific lab operations
2 Perception detect_labware, verify_state, calibrate Sensor and vision feedback
3 System recover, emergency_stop, wait_condition Error recovery and coordination

Drivers implement only the layers they support. A liquid handler (OT-2) implements Layer 1 actions but not Layer 0 motion primitives directly. A mobile manipulator might implement all four layers.

2. RobotDriver Protocol

Every robot driver must implement this async interface:

class RobotDriver(Protocol):
    async def connect(self, config: dict | None = None) -> bool
    async def disconnect(self) -> None
    async def execute(self, action: RobotAction) -> ActionResult
    async def stop(self) -> None
    async def get_state(self) -> RobotState
    def capabilities(self) -> list[ActionType]

Method Contracts

  • connect() — Initialize hardware or simulator. Returns True on success. Config dict is driver-specific (IP address, serial port, etc.).
  • disconnect() — Release resources. Idempotent — safe to call multiple times.
  • execute() — Run a single action. Must check connected state internally. Returns ActionResult with measurements, timing, and state snapshot.
  • stop() — Emergency halt. Must be fast (no cleanup, no graceful shutdown). After stop, driver is in disconnected state.
  • get_state() — Non-mutating query of current robot state.
  • capabilities() — Synchronous. Returns the ActionType values this driver handles.

3. ActionResult Contract

Every execute() call returns an ActionResult:

class ActionResult(BaseModel):
    success: bool
    status: ActionStatus          # COMPLETED | FAILED | CANCELLED
    action_type: ActionType
    measurements: dict[str, Any]  # e.g. {"volume_dispensed_ul": 100.0}
    state_after: dict[str, Any]   # robot state snapshot post-action
    error: str                    # REQUIRED when success=False
    duration_s: float             # wall-clock execution time

Key invariant: if success=False, error must be non-empty. This is enforced by a Pydantic model validator at construction time.

4. Safety Requirements

All robots default to safety_level: critical. This is non-negotiable.

SafetyVerdict is the shared data contract between safety checkers and the executor:

class SafetyVerdict(BaseModel):
    allowed: bool
    checker: str                  # which checker produced this
    reason: str                   # why it was blocked
    requires_confirmation: bool   # human-in-the-loop gate

Built-in safety checks:

  • WorkspaceBounds — rejects move actions outside defined 3D bounds
  • Future: force limits, collision detection, tip presence verification

Safety checkers are composable. The executor runs all applicable checkers before delegating to the driver. Any single allowed=False verdict blocks execution.

5. RobotManifest

Each robot directory contains a skill.yaml manifest:

name: "opentrons-ot2"
version: "0.1.0"
vendor: "Opentrons"
robot_category: "liquid-handling"    # liquid-handling | manipulation | transport | mobile-manipulator
safety_level: "critical"             # always critical for physical robots
robot_capabilities:
  degrees_of_freedom: 3
  repeatability_mm: 0.1
  end_effectors: [single_channel_p300]
  labware_types: [wellplate_96_flat, tiprack_300ul]

Parsed into RobotManifest (Pydantic model) for runtime validation.

6. Integration with LabClaw Orchestrator

lab-robot is one execution backend in the LabClaw ecosystem:

labclaw orchestrator
  ├── device-use    (GUI automation — reads instruments)
  ├── lab-robot     (physical execution — moves robots)
  └── lab-manager   (data + API layer)

Boundary rules:

  • device-use and lab-robot never import each other
  • Both register as tools with the orchestrator via capability descriptors
  • The orchestrator decides which backend handles each task
  • lab-robot actions are physical (pipette, move); device-use actions are digital (click, read)

Tool registration pattern:

# Orchestrator registers robot capabilities
capabilities = driver.capabilities()  # [ActionType.PIPETTE]
# Routes PipetteAction to lab-robot, ClickAction to device-use

7. Reference Implementation: Opentrons OT-2

The OT-2 driver (robots/opentrons_ot2/) is the reference PEI implementation.

  • Supports ActionType.PIPETTE (Layer 1)
  • Translates PEI PipetteAction into OT-2 protocol commands
  • simulate=True mode runs with no hardware (default for Phase 0)
  • Deck config via OT2DeckConfig Pydantic model
  • Protocol commands: pick_up_tip, aspirate, dispense, drop_tip

This driver demonstrates the pattern for all future robots:

  1. Parse skill.yaml into RobotManifest
  2. Implement RobotDriver protocol methods
  3. Translate PEI actions into vendor-specific commands
  4. Return rich ActionResult with measurements and state

8. Directory Convention

robots/
  _robot_template/     # copy this to start a new robot
    skill.yaml
    driver.py
    SOUL.md            # robot personality for AI context
    MEMORY.md          # calibration and usage history
    tests/
  opentrons_ot2/       # reference implementation
    skill.yaml
    driver.py
    models.py
    protocol_gen.py
    SOUL.md
    MEMORY.md
    tests/

Each robot is a self-contained package. The SOUL.md gives AI agents context about the robot's personality and quirks. MEMORY.md tracks calibration history.